  • Electric power systems designed for general, nonspecialized power distribution in which neither load conductor is directly grounded. These systems usually consist of a wall-mounted power distribution system fixture; they must include a device such as a line isolation monitor (LIM) or ground-fault detector to monitor the isolation degradation and alarms that typically operate at a limit between 2 to 5 mA. Isolated power systems are used in operating rooms and other anesthetizing locations and in \'wet\' locations; their installation must be performed in some of these places according to codes and standards. Periodic inspection and testing of these systems, including the function of the LIM, are also required.