?:abstract
|
-
OBJECTIVES The aim of this study was to evaluate the feasibility, procedural results, and 6-month outcomes of a novel transfemoral transcatheter mitral valve implantation (TMVI) system (Cephea) in patients with complex primary mitral regurgitation (MR). BACKGROUND TMVI is emerging as an alternative to surgery in patients with severe MR. To date, the great majority of TMVI systems use the transapical surgical approach. METHODS This study included consecutive patients undergoing transfemoral TMVI with the Cephea valve system. All patients were suboptimal candidates for catheter-based repair for anatomic reasons. Patients underwent clinical, echocardiographic, and computed tomographic angiographic follow-up at 6 months. Main outcomes were procedural success, peri-procedural complications, and valve hemodynamic status early and at 6-month follow-up. RESULTS Three patients with severe primary MR (2 women, mean age 79 ± 3 years) at prohibitive surgical risk (mean European System for Cardiac Operative Risk Evaluation II score 13.8 ± 2.4%) were included. The valves were successfully implanted in all patients, with no procedural complications. Post-procedural echocardiography showed normal valve function in all patients (mean transvalvular gradient 3 mm Hg; range: 2 to 4 mm Hg), no moderate to severe valvular or paravalvular leak, and no clinically significant left ventricular outflow tract obstruction. At 6 months, all patients had improved quality of life (mean improvement vs. baseline 16.4 ± 12 points). All valve hemodynamic parameters were maintained. Computed tomographic angiography confirmed annular stability, proper valve geometry and no structural failure. CONCLUSIONS The transfemoral delivery of a purposely designed mitral prosthesis (Cephea valve) was safe and feasible in prohibitive risk patients. Valve performance was sustained, and clinical outcomes improved at 6 months. Larger clinical studies are required.
|