Property | Value |
?:abstract
|
-
From 31.10. - 1.11. 2020 Slovakia has used the SD Biosensor Standard Q Ag-Test for nationwide tests for SARS-CoV-2, in which 3,625,332 persons from 79 counties were tested. Based on this data, we calculate that the specificity of the test is at least 99.6% (with a 97.5% one-sided lower confidence bound). Our analysis is based on a worst case approach in which all positives are assumed to be false positives. Therefore, the actual specificity is expected to exceed 99.6%.
|
is
?:annotates
of
|
|
?:creator
|
|
?:doi
|
-
10.1101/2020.12.08.20246090
|
?:doi
|
|
?:license
|
|
?:pdf_json_files
|
-
document_parses/pdf_json/edd274e4cd400bb9d588aebb42bfaebc331f6645.json
|
?:publication_isRelatedTo_Disease
|
|
?:sha_id
|
|
?:source
|
|
?:title
|
-
Analysis of the specificity of the SD Biosensor Standard Q Ag-Test based on Slovak mass testing data
|
?:type
|
|
?:year
|
|