PropertyValue
?:abstract
  • OBJECTIVE Real-time continuous glucose monitoring (rtCGM) in critically ill hospitalized patients holds promise; however, real-world data are needed. RESEARCH DESIGN AND METHODS We placed Dexcom G6 CGM on intensive care unit (ICU) patients at Montefiore Medical Center with confirmed coronavirus disease 2019 (COVID-19) infection and glycemic variability. We analyzed inpatient CGM accuracy using point-of-care (POC) glucose-CGM matched pairs and included patients for analysis regardless of clinical status. RESULTS We included 11 patients with CGM: 8 on continuous insulin infusion (CII), 8 on vasopressors, 8 intubated, 4 on high-dose glucocorticoids, 6 on renal replacement therapy, and 2 with anasarca. Accuracy was 12.58% for mean and 6.3% for median absolute relative difference. CGM reduced POC testing by ∼60% for patients on CII. CONCLUSIONS In this real-world preliminary analysis of rtCGM during critical illness, we demonstrate early feasibility, considerable accuracy, and meaningful reduction in the frequency of POC glucose testing.
?:creator
?:doi
  • 10.2337/dc20-2219
?:doi
?:journal
  • Diabetes_care
?:license
  • unk
?:pmid
?:pmid
  • 33361145
?:publication_isRelatedTo_Disease
?:source
  • Medline
?:title
  • Continuous Glucose Monitoring in the Intensive Care Unit During the COVID-19 Pandemic.
?:type
?:year
  • 2020-12-23

Metadata

Anon_0  
expand all